Tripp Lite
Tripp Lite SMART1200XLHG 1000 VA 750 W Medical UPS
Medical-grade tower UPS delivers 1000 VA / 750 W with four hospital-grade outlets and 10-minute full-load runtime
- Input Voltage Range120V
- VA Rating1000 VA
- Watts750 Watts
- Output VoltageLINE MODE: Sine wave line voltage 120V (-18% / +8%) BATTERY
- Outlets4
- Outlet TypeNEMA 5-15R-HG Hospital-Grade
- Medical-grade UL 60601-1 certification protects patient-care equipment
- 1000 VA line-interactive topology keeps output at 120 V from 81 to 143 V input
- Four hospital-grade NEMA 5-15R-HG outlets power critical devices safely
- 750 W capacity runs full load for 10 minutes and half load for 28 minutes
- RS-232 and USB ports enable remote monitoring and graceful shutdown
Medical-grade battery backup for patient-care
The Tripp Lite SMART1200XLHG is a 1000 VA line-interactive UPS tested to UL 60601-1 for use in patient point-of-care settings. It delivers 750 W through four hospital-grade NEMA 5-15R-HG outlets. The unit accepts 120 V input and maintains nominal output across an 81 to 143 V range.
Tower form factor with standard data ports
Place the tower on a stable surface and connect the included RS-232 or USB cable to a host computer for power-management software. The four HG outlets accept standard medical-grade plugs without adapters. Audible alarms signal power failure, overload and low battery with distinct beep patterns.
10 minutes of runtime at full 1000 VA load
At maximum rated load the internal VRLA batteries provide 10 minutes of backup; at half load the runtime extends to 28 minutes. Recharge to 90 % capacity takes 2 to 4 hours. The 480 J surge rating supplements the voltage-regulation circuitry.
Prednosti
- Medical-grade UL 60601-1 certification protects patient-care equipment
- 1000 VA line-interactive topology keeps output at 120 V from 81 to 143 V input
- Four hospital-grade NEMA 5-15R-HG outlets power critical devices safely
- 750 W capacity runs full load for 10 minutes and half load for 28 minutes
- RS-232 and USB ports enable remote monitoring and graceful shutdown
Špecifikácie
| Brand | TRIPP LITE |
|---|---|
| Series | SmartPro XL |
| Model | SMART1200XLHG |
| Input Voltage Range | 120V |
| VA Rating | 1000 VA |
| Watts | 750 Watts |
| Output Voltage | LINE MODE: Sine wave line voltage 120V (-18% / +8%) BATTERY MODE: PWM sine wave output 115V +/-5% |
| Outlets | 4 |
| Outlet Type | NEMA 5-15R-HG Hospital-Grade |
| Battery Type | Valve Regulated Lead Acid (VRLA) |
| Battery Run Time | Full load runtime: 10 minutes (1000 VA) Half load runtime: 28 minutes (500 VA) |
| Battery Recharge Time | 2 - 4 hour(s) to 90% |
| Alarms | 3 function audible alarm indicates power failure conditions (4 short beeps every 10 seconds), overload (continuous short beeps) and low battery (continuous tone) |
| Interface Port | RS-232, USB |
| Approvals | Tested to UL1778 (USA, Series AGSM1200PSR3HG), UL 60601-1 (Medical), cUL (Canada), Class A (Emissions) |
| Surge Energy Rating | 480 Joules |
| Features | Applications UL 60601-1 listed UPS provides ultimate safety and reliability for patient point of care solutions, telemedicine, nurse's stations, wireless communications systems, electronic prescribing applications, medical imaging and mission critical equipment. Not recommended for life support applications. Ideal UPS protection for applications in medical patient-care areas where UL 60601-1 listed medical electrical equipment is required Built-in isolation transformer with Faraday shield offers full isolation and removes common mode noise while maintaining full AC surge suppression |
| Dimensions | 12.50" x 7.50" x 9.50" |
| Weight | 45.20 lbs. |
| Prepravná hmotnosť | 21.80 kg |
| Rozmery balenia | 470 × 340 × 310 mm |
Otázky k tomuto produktu
How can I verify the SMART1200XLHG is working correctly after installation?
The unit runs a self-test on start-up and the three-function audible alarm confirms status: four short beeps every 10 seconds mean battery mode, continuous short beeps signal overload, and a continuous tone warns of low battery.
What ships with the SMART1200XLHG and what must I purchase separately?
The UPS includes a hospital-grade NEMA 5-15P input cord and four NEMA 5-15R-HG outlets; RS-232 and USB cables for monitoring are not included and must be bought separately.
Which capacity of the SmartPro XL medical UPS should I choose for my load?
This model delivers 1000 VA / 750 W with 10 minutes runtime at full load or 28 minutes at half load (500 VA); select it when your connected equipment stays within 750 W.
Aj v tomto oddelení