myVolts
myVolts RPD-60B 60W dual-output power supply 5V 12V 24V
Delivers 60 W dual output at 5 V and 12 V or 5 V and 24 V for medical devices
- TypeAT
- Medical-grade safety with 2×MOPP certification for patient-connected devices
- Dual 5V/12V outputs deliver 60W total power for versatile medical equipment
- 90-264VAC universal input enables worldwide deployment without voltage switching
- Extremely low leakage current under 150µA meets BF application requirements
- Compact 4×2 inch open-frame design fits space-constrained medical installations
Medical-grade dual-output supply
The myVolts RPD-60B is a 60W open-frame PCB power supply certified to 2 x MOPP under IEC/EN 60601-1. It accepts 90-264Vac input and provides regulated 5V and 12V or 5V and 24V rails for medical devices such as oral irrigators, hemodialysis machines, monitors and sleep-apnea units.
Replaces bulky linear or uncertified units
Engineers swapping out oversized linear supplies or non-medical brick adapters gain a 4 by 2 inch footprint with EMI class B compliance and leakage current below 150 µA. The jump is worth it when BF-rated patient contact and 2 x MOPP isolation are mandatory for regulatory approval.
Sustained load behaviour
Free-air convection cooling delivers continuous rated power without forced airflow. Short-circuit, overload and over-voltage protections latch off safely, while the green-design topology maintains efficiency across the full input range during 24-hour operation in clinical environments.
Points forts
- Medical-grade safety with 2×MOPP certification for patient-connected devices
- Dual 5V/12V outputs deliver 60W total power for versatile medical equipment
- 90-264VAC universal input enables worldwide deployment without voltage switching
- Extremely low leakage current under 150µA meets BF application requirements
- Compact 4×2 inch open-frame design fits space-constrained medical installations
Caractéristiques
| Type | AT |
|---|
Questions sur cet article
What are the 4" × 2" mounting dimensions for this open-frame PCB supply?
The unit uses a 4-inch by 2-inch footprint with mounting holes on that pattern, so it fits standard open-frame brackets or standoffs spaced to those centres.
How do I wire the dual outputs and what polarity mistake should I avoid?
Connect 5 V at up to 5.5 A and 12 V at up to 2.2 A to the marked terminals; the common error is reversing the 5 V and 12 V leads because both share the same ground reference.
When will the convection-cooled unit need its electrolytic capacitors refreshed?
Under continuous load the internal capacitors typically reach end-of-life after 30 years, so plan a recap around that interval if the supply runs 24/7.